| OUTPUTS |
| AC Current: |
0 to 10A range with the following
characteristics:- |
| |
No-Load Voltage (min.): 80V
Full-Load
Voltage (min.): 70V
Full-Load
Current (max.): 10A
Time:
2 hours
Safety:
Output transformer equipped with built-in temperature protection.
Primary side protected
by miniature circuit breakers. |
| 
|
0
to 25A range with the following characteristics:-
No-Load
Voltage (min.): 24V
Full-Load
Voltage (min.): 20V
Full-Load
Current (max.): 25A
Time:
1 hour
Safety: Output transformer equipped
with built-in temperature protection.
Primary side protected
by miniature circuit breakers. |
|
0
to 50A (short term max. 200A peak) range with the following characteristics:-
No-Load
Voltage (min.): 10V
Full-Load
Voltage (min.): 2V
Full-Load
Current (max.): 50A
Time:
1.5 minutes
Safety: Output transformer equipped
with built-in temperature protection.
Primary side protected
by miniature circuit breakers. |
| AC/DC
Voltage: 0 to 240V AC rated at 2A for 1 hour.
0 to
300V DC rated at 2A for 1 hour.
Auxiliary
Voltage: 240V AC rated at 0.5A.
240V
AC rated at 0.5A. 110V DC rated at 0.5A, 110V AC rated at 0.5A
& 30V DC rated at 1A.
If
the 110V DC & AC fixed outputs are used simultaneously the
rating is 0.2A each.
Safety (Voltage):
Re-triggerable fuses used in all voltage outputs |
| INPUTS |
|
| Digital: |
2 x wet contacts and 1 x dry
contact. |
|
|
| TIMER |
|
| Range: |
0.000 to 999.999 seconds, auto-ranging. |
| Resolution: |
1mSec. |
| Max. Testing Time: |
999.999 seconds. |
| MEASUREMENTS |
|
| Readout: |
Output
parameters are displayed on a large backlit LCD display.
Auto-ranging if output
is less than 10% of range. |
| Measured Quantities: |
AC
current output.
Timer. |
| Accuracy: |
±0.5% of metering. |
| AC Measurement: |
True RMS. |
| Set of Resistors: |
0.5Ω (10A max.), 1Ω
(5A max.), 25Ω (2A max.), 100Ω (0.7A max.) & 500Ω
(0.3A max.). |
| OPERATING VOLTAGES |
|
 |
| Prime Power: |
230 or 115V AC ±15%, 50/60Hz. |
| ENVIRONMENTAL |
|
| Operating
Temperature: |
-20ºC to +70ºC |
| Storage
Temperature: |
-40ºC to +70ºC |
| MECHANICAL DETAILS |
|
|
| Weight: |
10kg. |
| Carry Case: |
Reinforced aluminium, 13.5” x
9.5” x 6” (340mm x 240mm x 150mm). |
|
|
| DUE TO CONTINUING DEVELOPMENT
AND IMPROVEMENTS WEIS RESERVES THE RIGHT TO CHANGE THIS SPECIFICATION
WITHOUT NOTICE |